The Sapphire Multi-Therapy infusion system is a complete solution for varied clinical uses in hospital, ambulatory and EMS settings. Offering ease of use and reliability, Sapphire combines high-end volumetric infusion pump functionality with compact size and low weight.
Products Details
Sapphire is based on a unique combination of patented Flow Control Technology, innovative hardware design, intuitive touch-screen and software control, and low total cost of ownership.
With its robust design, easy maintenance, and built-in adaptability to evolving requirements, Sapphire delivers medication infusion confidence now, and far into the future. This same technology powers a wide range of dedicated infusion pumps, creating a single-platform that answers all medical infusion needs.
The Sapphire™ advanced multi-therapy infusion pump is designed for PCA, TPN, Continuous, Piggyback, Intermittent and Epidural treatments
Pump Specifications
- Pump Type: Volumetric, single-channel, touch screen pump, Eitan Medical Flow Control
- Multi-Therapy Infusion Modes: Continuous, Multi-step, TPN, Intermittent, PCA, Epidural, and Secondary (Piggyback)
- Dimensions: (H x W x D) 143 x 96 x 49 mm (5.63 x 3.78 x 1.93 in)
- Transportation: Approved for air and ground medical transportation**
- Weight: 418 g (14.7 oz) without battery
- Touch Screen: 3.2” color portrait QVGA (64K colors)
- Degree of Protection Against Ingress of Water and Dust: IP24 splash/dust standard according to IEC 60601-1-11
- Battery: Rechargeable Li-lon battery 7.4V 1960mA/h
- Battery Operation Time: ~24 hrs @ 125 mL/h
- Power Source: AC adaptor 100 – 240 VAC 50 – 60 Hz
- Volume (VTBI): 1 – 9,999 ml with increments of 0.1 mL
- Flow Rate: 0.1-999 mL/h with increments of 0.1 mL/h
- PCA Flow Rate: Basal rate 0.1-99.9 mL/h, Bolus rate up to 600 mL/h
- Epidural Flow Rate: Basal rate 0.1-25 mL/h, Bolus / Dose rate 125mL/h or 200 mL/h
- KVO Rate: Up to 20 mL/h with increments of 0.1 mL/h
- Flow Rate Accuracy: ± 2.5% according to IEC 60601-2-24
- Disposable Set: Automatic anti-free-flow set
- Downstream Occlusion: Up to 17.4 PSI (1.2 Bar)
- Prime: Manual or automatic priming (600 or 900 mL/h)