O2COMFORT - Comfortable Oxygen Therapy
O2COMFORT can administer oxygen in a traditional way, but also contains a comfort mode in which each dose of oxygen is administered only during the inhalation. The comfort settings 1 to 5 give an equivalent therapy of 1 to 5 litres per minute continuous flow (expressed in FiO2). The O2COMFORT flowmeter offers additional advantages over the traditional flowmeter and is highly suitable for every patient.
-
Most popular related searches
How does the comfort mode work?
The principle is simple: when in comfort mode, oxygen is delivered accurately during the first phase of the inhalation, which is the only phase that can be used for oxygen exchange in the capillaries of the lungs. The pneumatic system of the flowmeter doesn’t contain electronics or batteries and is triggered by the spontaneous breathing of the patient.
The continuous mode functions as a classical flowmeter.
Advantages
Premium patient comfort
Traditional, continuous oxygen therapy causes dehydration of the nose and the upper-airway mucosa that can lead to irritation, crust formation, nose bleeds and headache. The comfort mode keeps the moisture balance in the nose intact what stops the unpleasant side-effects for the patient.
Lower infection risk
The comfort mode allows humid air back into the patient’s nose. The need to use a humidifier disappears and so does the risk of an infection due to bacterial growth inside the humidifier.
Optimal oxygen consumption
The comfort mode ensures less oxygen waste compared to the continuous therapy while maintaining the same level of oxygen saturation. Long-term studies in different wards of the hospital validated a reduced oxygen consumption between 33% and 75%.
Higher therapy loyalty
It’s simple logic: when oxygen therapy no longer causes dehydration and stops being annoying, this likely makes the patient more loyal to the treatment. This higher therapy loyalty raises its effectiveness and eventually enhances the healing process.
Automatic closing
Interrupting treatment is as easy as taking out the nasal cannula. Comfort mode automatically cuts the oxygen flow but keeps the initial setting. When the patient re-inserts the nasal cannula the appropriate oxygen flow simply restarts.
Plug and play
The O2COMFORT flowmeter is a robust device that requires no maintenance. This easy-to-use instrument fits every type of connector and is fully certified. The life span is at least 8 years.
Technical Specifications
Inlet pressure: 5 bar (72.5 psi)
Comfort ?owrate: 1, 2, 3, 4 and 5 l/minequiv
Continuous ?owrate: Indicated from 0 to 15 l/min: 0 to 5 per half liter, 5 to 15 per 1 liter.
Up to 24 l/min ?ush ?ow
Oxygen use in
comfort mode: Position 1 = 0.260 l/min; position 2 = 0.420 l/min; position 3 =
0.560 l/min; position 4 = 0.700 l/min; position 5 = 0.860 l/min
Comfort mode
precision: Within ± 15% of the pulse intensity with each breath
Continuous mode
precision: From 0.5 to 5 ± 0.25 l/min; from 5 to 15 ± 0.5 l/min
Trigger method: Negative pressure during nasal inhalation
Breathing frequency: Up to 35 breaths per minute (constant minute volume)
Normal working
range: Temperature: 1.7°C to 40.6°C (35°F to 105°F);
Height: sea level to 3048 m (0 to 10,000 ft)
Deviant working
range: Effect on ?ow accuracy as a result of changes to ambient
temperature: +7.3% at 0°C (32°F) and -3.0% at +40°C (104°F)
Storage conditions: Temperature: -40°C to 60°C (-40°F to 140°F)
Maximum humidity: 95% non-condensing
Administering
requirement: A standard type nasal cannula for adults with a single lumen connection
Customer reviews
No reviews were found for O2COMFORT - Comfortable Oxygen Therapy. Be the first to review!